Hach FilterTrak 660 Laser Nephelometer

Hach FilterTrak 660 Laser Nephelometer  Hach Company
 

Overview

 

The new FilterTrak 660 Laser Nephelometer - the ultimate in simplicity and security in filtration management.

 

 

Finest Resolution in the Market

The Hach FilterTrak 660 sc Laser Nephelometer is designed specifically to detect changes in turbidity as low as 0.0005 NTU. Using advanced laser optics and signal processing, the instrument detects increased concentrations of submicron-sized particles that are a precursor to larger particles. This allows for early filter deterioration detection that meets or exceeds that of particle counters-all with the day-to-day convenience, simplicity, and reliability of a Hach Turbidity Meter. Operators can detect impending filter breakthrough, delineate filter ripening, and maximize effective filter run time.

 

Features/Benefits

 
  • Finest Resolution in the Market- The Hach FilterTrak 660 sc Laser Nephelometer is designed specifically to detect changes in turbidity as low as 0.0005 NTU.  Using advanced laser optics and signal processing, the instrument detects increased concentrations of submicron-sized particles that are a precursor to larger particles.  This allows for early filter deterioration detection that meets or exceeds that of particle counters--all with the day-to-day convenience, simplicity, and reliability of a Hach Turbidity Meter.  Operators can detect impending filter breakthrough, delineate filter ripening, and maximize effective filter run time.
  • Monitor, Optimize, and Report - USEPA method compliance means that treatment plants with conventional or membrane filtration can monitor and optimize, as well as report, using this single technology. 
  • Communications and Data Logging - Optimize data management with flexible direct digital communication options including 4-20 mA, digital MODBUS (RS-485 or RS-232), PROFIBUS DP, or Lonworks protocols.  Every sc100 and sc1000 controller may be equipped with wireless communication and a built-in data logger.
  • Simple, on-site calibration/verification - Routine maintenance on the FilterTrak 660 is simple. Unlike a particle counter, the FilterTrak 660 will not clog persistently. On-site calibration is fast and inexpensive, using Hach StablCal® 0.1 and 0.8 NTU standards.
  • Compatible with Hach Multi-Sensor, Multi-Parameter Digital Controllers - The FilterTrak sensor can be used with Hach's new sc100 or sc1000 Digital Controllers.  The sc100 controller accepts up to two sensors.  The sc1000 accepts up to eight sensors.  Multiple controllers can be networked to accommodate many more sensors and parameters, reducing the cost per measuring point.  Just plug in any Hach "plug and play" digital sensor and it's ready to use without software configuration.  "Plug and play" connectivity means there is no complicated wiring or set up.  Network the FilterTrak 660 sc Laser Nephelometer with any of Hach's digital sensors for measuring dissolved oxygen, pH, ORP, conductivity, and many other parameters.

Specifications

 
Approval Designed to comply with USEPA approved method 10133 for regulatory compliance reporting
Range: 0.000 to 5000 millinephelometric Turbidity Units (mNTU)
Note: 1000 mNTU = 1.000 NTU
Lowest Expected Reading: 5 mNTU using de-ionized, reverse osmosis water, based on statistical averages from three instruments
Accuracy: From 0 to 1000 mNTU: ±3% of reading or ±5 mNTU,
whichever is greater>

From 1000 to 5000 mNTU: ±2% of reading
Displayed Resolution: 0.001 mNTU up to 9.999 mNTU;
0.01 mNTU from 10.00 to 99.99 mNTU;
0.1 mNTU from 100.0 to 999.9 mNTU;
1 mNTU from 1000 to 5000 mNTU
Repeatability: Better than ±1.0% of reading at 24 mNTU; ±1.0% of reading at 800 mNTU; and ±1.0% of reading at 5000 mNTU as RSD (or as coefficient of variation)
Light Source: Class 1 laser product; with embedded 10 mW, 660 nm, Class 3B laser source (complies with 21 CFR 1040.10 and 1040.11. FDA Laser Accession No. 9911570)
Sample Flow Rate: 100 to 750 mL/min (1.6 to 11.9 gal/hour)
Recorder Output:* Two selectable for 0-20 mA or 4-20 mA; output span programmable over any portion of the 0 to 5500 mNTU range
Alarms:* Three set-point alarms, each equipped with an SPDT relay with unpowered contacts rated for 5A resistive load at 230 Vac
Power Requirements: 10.8 to 13.5 Vdc, 1.5 VA
Digital Communication: Network Card Compatible
MODBUS® RS-485, MODBUS® RS-232, LonWorks Protocol (optional)

Wireless Communication
IR Port to download into a handheld Personal Digital Assistant (PDA) or laptop computer via MODBUS®
 * Signal Output Module required.
Specifications subject to change.


For further information on this item, please download the following datasheets or contact Accurate Detection on 1300 887 735

Hach Filtertrak 660 Nephelometer Datasheet
Hach Filtertrak 660 Nephelometer Manual
Hach Intro to Laser Nephelometery
Hach Filtertrak 660 Case Study