Home  |  About Accurate Detection  |  Products  |  Latest News   |  Downloads   |  Contact Us
Useful links Site map 

Micro 100 Laboratory Turbidimeter

LABORATORY TURBIDIMETER

Designed to provide the accuracy demanded by today's scientist, the Laboratory Turbidimeter is ready to meet the challenge. Features like the Auto Alert Calibration Prompt and the Quick Connection Lamp Module, allow for simple calibration and low maintenance. With a resolution of 0.01 NTU at low turbidity readings and an extended range to 1,000 NTU, the is an ideal and affordable tool for research as well as routine analytical measurements.

STANDARD FEATURES:
Auto Ranging 0 - 1000 NTU
Instrument senses turbidity level of sample and automatically adjusts to the appropriate measurement.
Auto Alert Calibration Prompt Automatically prompts the operator when calibration of the instrument is needed.
Simple Calibration Procedures Calibration initiated with the push of a button ensures accurate readings.
RS-232 Output
Date, time, and NTU reading can be captured with a serial printer or data recorder.

Made in the USA

OPTIONAL FEATURES:
Infrared Light Source ( IR) Meets international standard ISO 7027 for turbidity measurement.
Pour-through assembly
Save time by pouring your grab sample through the measurement chamber.
Reusable Calibration Standards
Quick and easy calibration across all ranges without the need to mix Formazin
.


 

The comes with a complete set of calibration standards (0.02, 10 and 1000 NTU), two measuring cuvettes with light shield caps, operator's manual and power supply.

Note: The meets performance criteria specified by USEPA method 180.1 for NTU measurement and ISO7027 (IR version).

SPECIFICATIONS
Principal of Operation: Nephelometric non-ratio
Range: Auto ranging from 0 - 1000 NTU
Accuracy: ± 2% of reading plus 0.01 NTU
Resolution:

0.01 NTU (0 - 9.99 NTU)

0.1 NTU (10 - 99.9 NTU)

1 NTU (100 - 1000 NTU)

Response Time: < 6 seconds
Sample Size: 30 ml (27ml minimum)
Light Source: Quick connect Tungsten Filament or Infrared Operating
Temperature: 32° - 122°F (0° - 50°C)
Outputs: RS-232 Serial Port
Dimensions: 10.75 x 10 x 3.75 inches (27.3 x 25.4 x 9.5 cm)
Weight: 2.9 lbs. (1.32 kg)
Safety rating: UL, CSA, CE

For further information on this item, please contact us on 1300 887 735

Contact us: Australia sales@accuratedetection.com  Tel:1300 88 77 35
copyright © Accurate Detection 2003