Micro 1000 Laboratory Turbidimeter
|
INDUSTRIAL TURBIDIMETER
|
|
Designed to provide the accuracy demanded in today’s laboratory,
the
Ratio Turbidimeter is ready to meet the challenge. Features like
the Auto Alert Calibration Prompt and the Quick Connection Lamp
Module, allow for simple calibration and low maintenance. A full
time ratio (color correction) mode is also featured. With a
resolution of 0.0001 NTU at low turbidity readings and an
extended range to 10,000 NTU, the
is an ideal and affordable tool for research as well as routine
analytical measurements.
|
STANDARD FEATURES:
Extended Auto Ranging up to 10,000 NTU
Instrument senses turbidity level of sample and
automatically adjusts to the appropriate measurement.
Multiple Measurement Modes
NTU, EBC, and Nephelos, FNU, and FAU
Auto Alert Calibration Prompt
Automatically prompts the operator to initiate simple
menu-driven calibration.
Built-in Diagnostics
Full time diagnostic software automatically monitors and
detects errors.
Reusable Calibration Standards
Quick and easy calibration across all ranges without the
need to mix Formazin.
Ratio (Color Correction)
Automatic compensation for background color.
Made in the USA
OPTIONAL FEATURES:
Pour Through & Flow Through Assemblies
Save time by pouring your grab sample through the
measurement chamber or pressure feed using the flow
through assembly.
|
|

The
includes a complete set of Calibration Standards (0.02 - 1750
NTU), two measuring cuvettes with light shield caps, operator's
manual and plug-in power supply.
* Note: The
meets performance criteria specified ISO7027.
|
|
SPECIFICATIONS
|
|
Range: |
0 - 10,000 NTU
0 - 2,450 EBC
0 - 10,000 FNU
0 - 10,000 FAU |
|
Principal of Operation: |
Nephelometric |
|
Ratio (Color Correction): |
Full Time ON or OFF |
|
Accuracy: |
± 2% of reading plus 0.01 NTU (0 to 1000 NTU)
± 5% of reading (1000 to 4000 NTU)
± 10% of reading (4000 to 10,000 NTU) |
|
Resolution: |
0.0001 NTU on Lowest Range |
|
Response Time: |
less than 6 seconds |
|
Sample Size: |
30 ml |
|
Light Source: |
Quick connect Infrared |
|
Operating Temperature: |
32° - 122°F (0° - 50°C) |
|
Air Purge: |
Connection for external dry air supply |
|
Outputs: |
RS-232 Serial Port (Bidirectional) |
|
Shipping Dimensions: |
14 x 12 x 12 inches (35.5 x 30.4 x 30.5 cm) |
|
Weight: |
7.0 lbs. (3.18 kg) |
|
Certification: |
UL, CSA, CE, USEPA, ISO7027 |
|
For further information on this item, please call us on 1300 887 735
|